Instructions

HT9014 - HT9015
EN - 20
6.1.1. Safety
Comply with: IEC/EN 61010-1
Insulation: double insulation
Pollution degree: 2
Max height of use: 2000m (6562 ft)
Installation category: CAT IV 600V, CAT III 1000V to ground
6.1.2. General data
Mechanical characteristics
Dimensions (L x W x H): 215 x 74 x 43mm ; 8 x 3 x 2 in
Weight (including battery): 285g (10 ounces)
Max conductor size: 30mm (1in)
Supply
Battery type: 1 battery 9V NEDA 1604 IEC 6F22 JIS 006P.
Low battery indication:
” is displayed when the battery level is low
Battery life: about 200 hours
AutoPowerOff: after 15 minutes of idleness (not disabled)
Display
Characteristics: 4 LCD (max 6000 counts), decimal point, unit symbol
indication, bargraph and backlight
Sample rate: 2 times/sec
Conversion mode: TRMS
6.2. ENVIRONMENTAL CONDITIONS
6.2.1. Climatic conditions
Reference temperature: 18°C ÷ 28°C (65°F ÷ 83°F)
Operating temperature: 5 ÷ 40 °C (41°F ÷ 104°F)
Operating humidity: <80%RH
Storage temperature: -20 ÷ 60 °C (-4°F ÷ 140°F)
Storage humidity: <80% RH
This product conforms to the prescriptions of the European directive on low voltage
2006/95/EEC and to EMC directive 2004/108/EEC
6.3. STANDARD ACCESSORIES
Test leads – Code 4413-2
Adapter + type K bead probe
Carrying bag
Battery
User manual
6.4. OPTIONAL ACCESSORIES
The herewith type K probes are avaliable:
Model Description
Temperature
range
Accuracy
(at 100°C)
Probe length
(mm)
Probe diameter
(mm)
TK107
Air and gas
temperature
-40 ÷ 800 °C
± 2.2rdg 200 1.5
TK108
Inner temperature of
semisolid
substances
-40 ÷ 800 °C
± 2.2rdg 200 3
TK109
Inner temperature of
fluids
-40 ÷ 800 °C
± 2.2rdg 200 4
TK110
Surface temperature
-40 ÷ 400 °C
± 2.2rdg 200 5
TK111
Surface temperature
with fixed tip at 90°
-40 ÷ 400 °C
± 2.2rdg 260 5