User's Manual HTC Manometer Please read this user's manual thoroughly before using this unit and keep it properly for your future reference.
Contents 1. Safety........................................................................................... 1 2. Introduction ............................................................................... 2 3. Error Warning ............................................................................ 2 4. Specification .............................................................................. 3 5. Meter Description .................................................................... 5 6. LCD display .
1. Safety Read the following safety information carefully before attempting to operate or service the meter. Use the meter only as specified in this manual; otherwise, the protection provided by the meter may be impaired. ▲ Environment conditions: 1) RH≤90%(Non-Condensation) 2) Operating Temperature-0~40℃ Maintenance and cleaning 1) Repairs or servicing not covered in this manual should only be performed by qualified personnel. 2) Wipe the unit with a dry soft cloth.
2. Introduction This product is aimed to measure different pressure values, here are some units for Choice: inH2O, psi ,bar, mbar , kPa, inHg, mmHg ,ozin2 ,,ftH2O, cmH2O ,kgcm2.equipped with simple Tube connection. It is widely used in measurement of ventilation and air conditioning system, clean room or any gas pressure system troubleshooting .Easy operation and clear reading. 3. Error Warning When internal diagnostic test fails, an error message will appear on the LCD display.
4. Specification PM-6102 PM-6105 PM-6115 unit Range Resolution Range Resolution Range Resolution inH2O ±55.40 0.01 ±138 0.1 ±415.2 0.1 Psi ±2.000 0.001 ±5.000 0.001 ±15.00 0.01 Mbar ±137.8 0.1 ±344.7 0.1 ±1034 1 KPa ±13.79 0.01 ±34.47 0.01 ±103.4 0.1 inHg ±0.407 0.001 ±1.018 0.001 ±30.54 0.01 mmHg ±103.4 0.1 ±258.5 0.1 ±775.7 0.1 ozin² ±32.00 0..01 ±80 1 ±240.0 0.1 ftH2O ±4.612 0.001 ±11.53 0.01 ±34.60 0.01 cmH2O ±140.6 0.1 ±351.5 0.
unit PM-6130 Range inH2O Psi PM-6175 Resolution Resolution Range ±830 1 ±2078 1 ±30.00 0.01 ±75.00 0.01 Mbar ±2608 1 ±5168 1 KPa ±206.8 0.1 ±517.1 0.1 inHg ±61.08 0.01 ±15.00 0.01 mmHg ±1551 1 ±3878 1 ozin² ±480.0 0.1 ±1200 0.1 ftH2O ±69.2 0.01 ±173.0 0.01 cmH2O ±2109 1 ±5272 1 kgcm² ±2.109 0.001 ±5.250 0.001 bar ±2.068 0.001 ±5.138 0.001 Maximum Pressure Psi 100Psi accuracy ±0.3%FS Linearity/Hysteresis ±0.1%FS Repeatability ±0.
5. Meter Description ① Air intake ② LCD Display ③ Power key ④ Backlight key ⑤ Unit key ⑥ Hold button ⑦ record key 6.
7. Operation steps 7.1 Open battery cover and install a 9-volt battery in the battery compartment. Press power key to start measuring 7.2 press UNITS key,choosing proper unit 7.3 connect Tube 7.4 read the testing values on LCD display Note : when the meter is on, no gas pressure input, the meter will not necessarily be zero, you need to reset by yourself. 8. Buttons function 8.
“ ”for three seconds,the meter exits record model. 8.4 key;after power on the meter,press“ ”,it can convert the units of measurement:bar,mmHg,Ozin2,Kgcm2,psi,inH2O,Kpa,ft H2O,inHg,cmH2O,mbar. 8.5 key; 8.5.1 After the meter power on,press“ be locked, press“ ”button once,the screen will ”again,it will cancel this function. 8.5.2 After power on the meter,no gas pressure input,press“ ”for three seconds,LCD screen display“----”,the meter reset. 8.
9. Notes 1. Do not use it in high temperature and humidity。 2. Remove the battery when the meter is to be stored for long periods of time to avoid battery leakage. 3. Replace new batteries when the low battery icon " 10. Accessories ① Manual ② Canvas ba ③ 6F22 9V battery ④ ∮4.5mm mental gas mouth ⑤ ∮5.0mm*30mm soft Tube 2PCS 2 PCS ⑥ ∮5.0mm*30mm hard Tube 2PCS www.htcinstrument.com 8 " appears.