Error Scenarios Icon Possible Causes Solution
Post-exercise
measurement
You performed moderate-to-
high intensity exercise before
the measurement.
Rest for at least half
an hour after
exercise, then
measure your blood
pressure.
Table2-2 Error notications during a measurement
Error Scenarios Possible Causes Solution
Poor signals During the measurement,
the watch pressed your
chest, your body moved, or
you breathed deeply.
During the measurement, avoid
pressing your chest with the watch,
keep your body stable, and breathe
evenly.
Not being still You moved your arms or
ngers, or spoke during the
measurement.
Keep still. Do not speak or move your
arms or ngers.
Irregular
heartbeat
Irregular heartbeats were
detected during the
measurement.
Take a break of one to two minutes
and perform the measurement again.
If the issue persists multiple times,
contact a doctor.
Airbag ination
error
The airbag was installed
incorrectly or damaged.
Check the airbag and ensure that it is
installed correctly. Check whether the
airbag is damaged. If so, replace the
airbag in a timely manner.
Worn incorrectly You did not wear your
watch correctly during the
measurement.
Wear your watch correctly based on
the instructions.
Low battery Your watch's battery is too
low.
Charge your watch to ensure that the
battery level is sucient.
No airbag There is not an airbag in the
watch.
Install a suitable airbag and strap
based on your wrist circumstance.
Blood pressure management
14