WATCH D User Guide-(MLY-B10,01,en-gb)

Table Of Contents
Error Scenarios Possible Causes Solution
Unsuitable
airbag
The airbag that is installed
does not match the set wrist
circumference: The airbag is
incorrectly installed or the
wrist circumference is
incorrectly set in the Blood
pressure app.
Measure your wrist circumference as
instructed in the Quick Start Guide,
install a suitable strap and airbag
accordingly, and go to Blood pressure
> Settings > Wrist circumference to
select the correct wrist circumference.
Abnormal
deation
The airbag deated slowly
after the measurement was
complete.
Take o your watch immediately and
clean the air inlet and outlet based on
the solution for handling slow
deation.
Takes a long
time to inate
The ination took too long
during the measurement.
Take o your watch immediately, wait
until the airbag is fully deated, and
start a measurement again. If the
fault occurs multiple times, contact a
service centre for it to be repaired.
High pressure The pressure was too high
during the measurement.
Do not press the airbag or strap
during the measurement.
Solution for handling slow deation:
Perform the following steps to clean the air inlet and outlet. If the issue persists,
contact a service centre for it to be repaired.
a Remove the short strap (including the buckle) and keep the long strap and airbag.
Keep the bottom cover facing up and unfold the long strap and airbag in your
hand.
b Dip a cotton swab in an appropriate amount of detergent and water. Use the cotton
swab to slowly wipe the air lter back and forth 20 times along the shorter side of
the watch body until the air lter is clean. Do not exert too much force, otherwise
the air lter may fall o.
c Wipe the long strap and airbag dry and put them to one side for half an hour.
Managing blood pressure
1 On the home screen of the Huawei Health app, touch the Blood pressure card. If there is
no blood pressure card, touch EDIT to add it.
2 Touch Blood pressure management plan > Go, and follow the onscreen instructions to
enter basic information and join the plan.
Other settings
Swipe up on the measurement result screen
Blood pressure management