Figure 2. 4.1 
Volume – This is the total volume that will be primed through the system. Enter a 
value between 120 and 9999uL. 
Speed  – This adjusts the percentage of the pumps maximum speed. More viscous 
liquids will require a slower pump speed to avoid cavitation. Do not exceed more 
than 30% speed on any value over 1000ul. 
Tip Touch – Activate or deactivate tip touching after each dispense. 
Tip Touch – Adjust where in the Priming trough you would like to touch the tips 
after a prime. This is helpful when running viscous liquids. 
2.5 - Empty Dispenser 
This feature reverses the pump, removing most of the liquid from the manifold 