This unit has been calibrated and is ready for use as received. After set-up, calibration is recommended. Calibration Check: Verify the procedure to be used and the accuracy of the volume indicator by using the apparatus to measure containers or molds of determinable volume that dimensionally simulate test holes that will be used in the field. The apparatus and procedure shall be such that these volumes will be measured to within 1.0 percent.
on the volume indicator. Note and record the pressure. Depending upon the type of apparatus, this pressure may be as high as 5 psi (34.5 kPA). It will usually be necessary to add load (surcharge) to the apparatus to prevent it from rising (Note 3). Note and record the total amount of load added. The difference between the initial and the final readings of the volume indicator is the indicator volume value for the container.
Table 1: Volume of water per gram based on Temperature °C °F Water Volume ml/g 12 53.6 1.00048 14 57.2 1.00073 16 60.6 1.00103 18 64.4 1.00138 20 68 1.00177 22 71.6 1.00221 24 75.2 1.00268 26 78.8 1.0032 28 82.4 1.00375 30 86 1.00435 32 89.6 1.00497 Values other than show may be obtained by referring to: Handbook of Chemistry and Physics Chemical Rubber Publishing Co.
Warranty Humboldt Mfg. Co. warrants its products to be free from defects in material or workmanship. The exclusive remedy for this warranty is Humboldt Mfg. Co., factory replacement of any part or parts of such product, for the warranty of this product please refer to Humboldt Mfg. Co. catalog on Terms and Conditions of Sale. The purchaser is responsible for the transportation charges. Humboldt Mfg. Co.