2.09 product manual H-4374 & H-4374.
Introduction The motorized sand equivalent shaker features two separate electronic timers (one preset at 45 seconds and the other preset at 10 minutes) to increase accuracy and ease of operation compared to a spring-type timer. It is recommended for samples in the laboratory. The consistent, repeatable oscillation of the apparatus eliminates operatorcaused variations.
Operation Control Panel: The controls are situated on the front, right hand side of the machine base. Preset Travel and Speed: The shaker has a throw of 8 ± 0.04” and operates at 175 ± 2 cycles per minute. Method: Prepare the sample to be tested according to the governing agency standard(s). Place the clear plastic cylinder containing the test sample and rubber stopper in position on the slide assembly of the shaker.
Warranty Humboldt Mfg. Co. warrants its products to be free from defects in material or workmanship. The exclusive remedy for this warranty is Humboldt Mfg. Co., factory replacement of any part or parts of such product, for the warranty of this product please refer to Humboldt Mfg. Co. catalog on Terms and Conditions of Sale. The purchaser is responsible for the transportation charges. Humboldt Mfg. Co.