Press (INCREASE)or (DECREASE) repeatably to scroll through
the available channel names for an appropriate choice. Choices for channel
namesare:CH1,CH2,CH3,CH4,LOAD1,LOAD2,LOAD3,LOAD4,DISP1,
DISP2,DISP3,DISP4,FLOW,STAB,VERT,HORZ,PRES1,PRES2,PRES3,PRES4,
VOLU1,VOLU2,VOLU3andVOLU4.
NOTE: It is important to assign an appropriate name when naming
channels because the HM-3000 will choose a unit of measure based on the
name assigned. For example, if you are configuring a load cell, name the
channelLoad1,2,3or4.Thiswillensurethatthechannelreadsinlbs(kN).
ThesameappliestoDisplacement(DISP1,2,3or4)readingininch(mm)
andPressure(PRES1,2,3or4)readinginpsi(kPa)etc.
OnceanamehasbeenchosenforachannelPress to return to the
previous screen. You should see this screen (LOAD1 was chosen as name):
Sensor Min
TheSensorZero(SENSORMIN)isnotcongurableandissetat0.
Sensor Max
Press
(SENSORMAX)tosettheSensorMaximumvalue.
You should see this screen:
“CH1” CONFIGURATION
* NAME = “CH1”
* INCREASE
* DECREASE
CH1 CONFIGURATION
* NAME = “LOAD1”
* SENSOR MIN = 0
* SENSOR MAX = “VALUE”
18
“CH1” MAX SETUP
* SENSOR MAX = “000000”
* INCREASE
* DECREASE