Manual
28
Peakvaluesareshownonthersttwolines.Toreviewallthedata,
press
(REVIEWDATA).Youshouldseethisscreen:
Pressing (PRINTDATA)willprintthedatatoaprinterorsaveale
in ASCII mode to your computer using Humboldt Material testing Software
(HMTS)orWindowsHyperterminal.Oncethedatahasbeenprinted
or transferred, you can begin another test after you have loaded another
sample and raised the head to the start position, close to the load cell. by
pressing
(RUNMARSHALLTEST).
If you don’t have a printer or a computer connected to the HM-3000, you
can press
(SCROLL DATA) to scroll the data. You should see this
screen:
Using the
(NEXT)keytoscrollthroughthereadingsandthe
(LAST) key to view the last reading taken.
To return to any prior screen you can always press
USER DEFINED
A user defined test could be used for any other test beside Marshall and
CBR / LBR. The test would require up to 3 channels associated with it.
User Defined test such as Triaxial (CU) would require 3 channels (load,
displacement, pore water pressure), A Triaxial (UU) requires 2 channels
(load and displacement), Unconfined Compressive strength would require
2 channels (load and displacement), Soil-Cement and TSR would requires a
single channel (load).
USERDEFINED"UnconnedCompressivestrengthtest"Centerthesoil
sampleontheplaten.Pressthe
button to raise the sample to the
start position, close to the unconfined upper platen. There should be about
a 1/8" (3.000mm) gap between the top of the sample and the unconfined
upper platen.
* PRINT DATA
* SCROLL DATA
* RUN MARSHALL TEST
STAB “READING”
FLOW “READING”
* NEXT
* LAST Time = 0.0 sec