8.OO troubleshooting
8.O1 Troubleshooting Table
Problem Cause Solution
1. Dead Unit - Power LED
not illuminated
a. No incoming main electrical power
Check wiring from dish machine
Check for power at dish machine connection
b. Disconnected cable between DMx main
board and power supply
Have a qualified technician open the cabinet and restore
the connection
c. Damaged power supply or DMx main board
Replace faulty components or units as needed
2. No detergent feed a. Unit is in Standby mode
Ensure the unit is not in Standby mode (see page 20)
b. No detergent signal is being received
Check signal wiring to the dish machine.
Check that the dish machine is sending a signal
c. An incorrect signal
Check DMx settings
d. Conductivity reading above setpoint value
(when in probe mode)
Check conductivity probe setpoint value
Clean or replace conductivity probe
e. Disconnected cable between DMx main
board and power supply
Have a qualified technician open the cabinet and restore
the connection
f. Product drum is depleted
Check the system for Out of Product or Low Detergent
Alarms
Physically check and replenish empty containers
g. Damaged power supply or DMx main board
Replace faulty components or units as needed
3. Excessive detergent
usage
a. Deposits on Conductivity probe (when in
probe mode) or faulty probe
Check signal wiring to the dish machine
Check that the dish machine is sending a signal
b. Wash tank drain stuck open
Clear wash tank drain obstruction
c. Wash tank, tubing, or fittings are leaking
Eliminate leaks in wash tank or delivery tubing/fittings
4. No rinse or sanitizer feed a. Unit is in Standby mode
Ensure the unit is not in Standby mode (see page 20)
b. No rinse signal is being received
Check signal wiring to the dish machine
Check that the dish machine is sending a signal
c. Detergent signal did not occur within 90
seconds before rinse signal (when in Door
mode)
Check detergent signal wiring
Check DMx settings (e.g. Door Mode and Rinse on
Rinse)
d. Disconnected cable between DMx main
board and power supply
Have a qualified technician open the cabinet and restore
the connection
e. Product drum is depleted
Check the system for Out of Product or Low Detergent
Alarms
Physically check and replenish empty containers
f. Faulty DMx main board
Replace DMx Main board
5. Insufficient product
usage/delivery
a. Squeeze tube is worn and needs
replacement
Replace squeeze tubes as needed
b. Product drum is depleted
Check the system for Out of Product or Low Detergent
Alarms
Physically check and replenish empty containers
6. DMx mobile application
cannot login
a. User credentials are invalid or not set up
correctly
Check user login and password and retry
Contact your Chemical Company Admin or Hydro
Technical Support to set up the user correctly
g. User is in an area without a connection AND
does not have a valid token to work offline
Go to an area with a cellular or Wi-Fi connection and log-
in to get a validated token that allows you to work offline
WARNING: Electrical installation and maintenance should be completed by a qualified electrician. All local and
national electrical regulations are to be observed.