User's Manual
Relative
Humidity Range
45%
to 75% non-condensing
Maximum
Operating Altitude
2000
m above sea-level
Storage and Transportation
Temperature
Range
-
10 to +40 °C
Relative
Humidity Range
5%
to 95% non-condensing
PCR Performance
Thermal Ramp
Rate
Pre
Heater
Maximum Ramp
Rate 1.1°C per second
Thermal Uniformity
1.
Cup Holder Heater (23->98°C):
Uniformity achieved
after 1.5~5 minutes Maximum
±
0.1°C steady-state error
2.
Disk High Voltage Heater (50->99°C):
Uniformity achieved
after 24~33
seconds Maximum
±
1.2°C steady-state error
3.
Disk Low Voltage Heater (50->59°C):
Uniformity achieved
after 18~27
seconds Maximum
±
1.0°C steady-state error
Fluorescence Detection
LED
illumination
Compact
camera module for image capture
Operating Environment
Temperature
Range
+22
to +28 °C (Ambient temperatures above 25°C
can reduce the
rate of cooling)