Data Sheet
Switzerland
www.ideal-tek.com
ACCURACY SPECIFICATIONS
RESISTANCE, IMPEDANCE
RANGE RESOLUTION 100 Hz 1 kHz 10kHz
1 Ω 0.001 Ω 0.7% + 50 0.7% + 50 0.7% + 50
10 Ω 0.01 Ω 0.7% + 8 0.7% + 8 0.7% + 8
100 Ω 0.01 Ω 0.2% + 3 0.2% + 3 0.2% + 3
1000 Ω 0.1 Ω 0.2% + 3 0.2% + 3 0.2% + 3
10 kΩ 0.001 kΩ 0.2% + 3 0.2% + 3 0.2% + 3
100 kΩ 0.01 kΩ 0.5% + 5 0.5% + 5 0.5% + 5
1000 kΩ 0.1 kΩ 0.5% + 5 0.5% + 5 0.5% + 5
10 MΩ 0.001 MΩ 2.0% + 8 2.0% + 8 5.0% + 8
Accuracy for the ranges 1 R ~ 100 R is specied after subtract of the offset resistance.
CAPACITANCE
RANGE RESOLUTION 100 Hz 120 Hz 1 kHz 10 kHz
1000 μF 0.1 μF 0.5% + 5 0.5% + 5 NA NA
100 μF 0.01 μF NA 0.3% + 3 0.5% + 5 NA
10 μF 0.001 μF NA 0.2% + 3 0.2% + 3 0.5% + 5
1 μF 0.1 nF NA 0.2% + 3 0.2% + 3 0.2% + 3
100 nF 0.01 nF NA 0.2% + 3 0.2% + 3 0.5% + 3
10 nF 0.001 nF NA 0.5% + 5 0.2% + 3 0.5% + 3
1000 pF 0.1 pF NA NA 0.5% + 5 0.5% + 3
100 pF 0.01 pF NA NA 0.5% + 10 0.8% + 20
10 pF 0.001 pF NA NA NA 1.0% + 50
Accuracy for the ranges of 10 pF~1000 pF is specied after subtract of stray capacitances for test leads.
INDUCTANCE
RANGE RESOLUTION 100 Hz 1 kHz 10 kHz
10 μH 0.001 μH NA NA 1.0% + 5
100 μH 0.01μH NA 1.0% + 5 0.7% + 3
1 mH 0.1 μH 0.7% + 10 0.5% + 3 0.5% + 3
10 mH 0.001 mH 0.5% + 3 0.2% + 3 0.5% + 3
100 mH 0.01 mH 0.5% + 3 0.2% + 3 NA
1 H 0.1 mH 0.2% + 3 NA NA
ANTI-MAGNETIC/ANTI-ACID GOLD PLATED STAINLESS STEEL TIPS
ST5-IT-T
Standard replaceable tips
ST5-IT-TSHP
High precision sharp
replaceable tips
ST5-IT-TB
High precision bent
replaceable tips