PERFORMANCE VERIFICATIONS
Perform the following analysis; if the meter conforms to the limits listed in Table 1 through 7 the meter is
functioning correctly. If the meter does not conform to any of the listed limits the calibration procedure
must be performed.
Performance Verification Preparation
1. Turn on the calibrator, allow calibrator to warm up. Temperature stabilization
should be reached after 30 minutes.
2. Remove battery cover and using a calibrated meter to ensure the battery measures a minimum of 7.5V
DC. If the battery measures under 7.5V DC, replace the battery (see Battery Replacement page 11)
before beginning the performance test.
3. Input the values listed in Table 1 through 7.
Table 1 AC Voltage Test
Function
/Range
Input
Low
Limit
High
Limit
Model Number
V AC 750V
120V AC @ 45Hz
118.3
121.7
61-763, 61-765
V AC 750V
120V AC @ 400Hz
112.3
127.7
61-763, 61-765
V AC 750V
700V AC @ 45Hz
689.0
711.0
61-763, 61-765
V AC 750V
700V AC @ 400Hz
657.5
742.5
61-763, 61-765
Table 2 DC Voltage Test
Function
/Range
Input
Low
Limit
High
Limit
Model Number
V DC 1000V
350V
346.0
354.0
61-763, 61-765
V DC 1000V
900V
886.0
914.0
61-763, 61-765
Table 3 AC Current Test
Function
/Range
Input
Low
Limit
High
Limit
Model Number
A AC 660A
10A AC @ 50Hz
9.3
10.7
61-763, 61-765
A AC 660A
10A AC @ 400Hz
8.9
11.1
61-763, 61-765
A AC 660A
500A AC @ 50Hz
489.5
510.5
61-763, 61-765
A AC 660A
500A AC @ 400Hz
469.5
530.5
61-763, 61-765
Table 4 DC Current Test
Function
/Range
Input
Low
Limit
High
Limit
Model Number
A DC 660A
100A DC
97.5
102.5
61-765
A DC 660A
500A DC
489.5
510.5
61-765
Table 5 Resistance Test
Function /Range
Input
Low
Limit
High
Limit
Model Number
1000
100
98.5
101.5
61-763, 61-765
10000
1K
985
1015
61-763, 61-765