Ranges & Accuracies
AC Converter:61-795modelisaveragingsensing,rmscalibrated
Accuracy:Accuracyisspeciedas+/-(apercentageofthereading+axed
amount)at23°C±5°C(73.4°F±9°F),lessthan75%relativehumidity.
Temperature Coefficient:0.1timestheapplicableaccuracyspecication
from32°Fto64°Fand82°Fto122°F(0°Cto18°C;28°Cto50°C).
*Continuitytestcurrent>200mA@R<5Ωandopencircuitvoltage(4-24VDC)
perEN61557-2/VDE0413Part4.
**AutoRanging.Opencircuitvoltage:<1.3xV
0
.
Accuracy of nominal voltage:
0%to+10%.Shortcircuitcurrent:<3.0mA.Nominaltestingcurrent:1mA@
1kΩ x V(1mA@500kΩ)
Function Range Accuracy
Overload
Protection
DCVoltage 0.1-600.0V ±(0.5%+1) 600VACRMSmax
ACVoltage 0.1-600.0V ±(0.8%+4) 600VACRMSmax
Resistance 0.1-199.9Ω ±(2.0%+3)
600VACRMSmax
for 1 minute
Continuity* 0.01-19.99Ω ±(2.0%+3) 600VRMSmax
Insulation
Resistance**
250V
0.001-0.100MΩ ±10
600VRMSmax
0.101-3.999MΩ
±(2.0%+5)
4.00-39.99MΩ
40.0-399.9MΩ
±(5.0%+5)
400-1000MΩ
500V
0.001-0.250MΩ ±15
0.251-3.999MΩ
±(2.0%+5)4.00-39.99MΩ
40.0-399.9MΩ
400-2000MΩ ±(5.0%+5)
1000V
0.001-0.250MΩ ±15
0.251-3.999MΩ
±(2.0%+5)4.00-39.99MΩ
40.0-399.9MΩ
400-1000MΩ ±(3.0%+5)
1000-4000MΩ ±(5.0%+10)