Insulation Resistance
1. Press to turn on the instrument.
2.Presstoselect1000V,500V,or250Vtestvoltage.
3.Insertthetestleadsinthecorrespondinginputterminalsofthe
instrument.
4. Disconnect the circuit under test from power and isolate all eventual
loads.
5.Applyleadstowiringundertest.
6.Setthetimerperiodfortesting.
• Defaulttimeis30seconds
• Holdfor1s.Then,usetoselecttime:1min.to10min.
Press to confirm selection.
• Tocontinuoustest:Pressand.Lockisdisplayed
and unit beeps. Press again to stop continuous mode.
7. Press to start the insulation testing. The resulting megohm
readingisdisplayed.Keeptheleadsonthetestpointstoallowthe
circuit to discharge.
Warning: The symbol on the display means that the
instrument is charging or discharging the circuit and indicates a
shockhazardispresent.
61-795
Digital Insulation Tester
1000V
Test
M
500V
M
250V
M
Red
Black
P
N
PE
R
E
R
T
TEST
Battery & Fuse Replacement
WARNING: To avoid electric shock, disconnect test leads before
removing battery cover.
WARNING: For continued protection against fire, replace only
with fuse of the specified voltage, current and rupture speed ratings.