Instructions / Assembly

Page 3
PERFORMANCE VERIFICATIONS
Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning
correctly. If the meter does not conform to any of the listed limits, the calibration procedure must be
performed.
Check the fuse as a BBS, 1A/600V, 10.4*35 (0.41x1.38 inch) UL/CSA, interrupting rating 10kA.
Check case, leads, display, etc, for any defects.
Table 1 Performance Verification: Model 61-795
Function Setting
/Range
Apply Reading Specification
DCV 550V DC 547.2 to 552.8 [±0.5% ±1 digits]
ACV 550V AC 545.2 to 554.8 [±0.8% ±4 digits]
Continuity /
<30 Buzzer sounds Not specified
Resistance:
LO Simulated leads .08 to .12 [±0.02 + 0.05 per lead]
LO 0 -0.03 to +0.03 [±0.03]
LO 5 4.87 to 5.13 [±2.0% ±3 digits]
LO 19 18.59 to 19.41 [±2.0% ±3 digits]
/
190 185.9 to 194.1 [±2.0% ±3 digits]
LO 10 20.2mA to 25.8mA Source check
LO 1 202mA to 258mA Source check
Insulation Resistance:
M, 250V 10M 250 to 275Vdc Source Check
M, 500V 10M 500 to 550Vdc Source Check
M 1000V 10M 1000 to 1100Vdc Source Check
0 0.001 to 0.010 ±10 digits
100k 0.090 to 0.110 ±10 digits
1M 0.975 to 1.025 [±2.0% ±5 digits]
10M 9.75 to 10.25 [±2.0% ±5 digits]
M, 250V
100M 94.5 to 105.5 [±5.0% ±5 digits]
100k 0.085 to 0.115 ±15 digits
1M 0.975 to 1.025 [±2.0% ±5 digits]
10M 9.75 to 10.25 [±2.0% ±5 digits]
100M 97.5 to 102.5 [±2.0% ±5 digits]
M, 500V
1000M 945 to 1055 [±5.0% ±5 digits]
100k 0.085 to 0.015 ±15 digits
1M 0.975 to 1.025 [±2.0% ±5 digits]
10M 9.75 to 10.25 [±2.0% ±5 digits]
100M 97.5 to 102.5 [±2.0% ±5 digits]
M, 1000V
1000M 965 to 1035 [±3.0% ±5 digits]
M, 500V 500K <500V Source Check
Check on each
Voltage range
0M <3.0mA Source Check
Form number TM61795 Rev 3 Jun 2008