PERFORMANCE VERIFICATIONS
Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning
correctly. If the meter does not conform to any of the listed limits, the calibration procedure must be
performed.
Check the fuse as a BBS, 1A/600V, 10.4*35 (0.41x1.38 inch) UL/CSA, interrupting rating 10kA.
Check case, leads, display, etc, for any defects.
Table 1 Performance Verification: Model 61-795
Function Setting
/Range
Apply
Specification
DCV
550V DC
0.5% ±1 digits]
ACV
550V AC
0.8% ±4 digits]
Continuity Ω/
<30Ω
Not specified
Resistance:
LOΩ
Simulated leads
0.02Ω + 0.05 per lead]
LOΩ
0.03Ω]
LOΩ
5Ω
2.0% ±3 digits]
LOΩ
19Ω
2.0% ±3 digits]
Ω/
190Ω
2.0% ±3 digits]
Insulation Resistance:
MΩ, 250V
10MΩ
Source Check
MΩ, 500V
10MΩ
Source Check
MΩ 1000V
10
Source Check
MΩ, 250V
0Ω
±10 digits
100kΩ
±10 digits
1MΩ
2.0% ±5 digits]
10
2.0% ±5 digits]
100MΩ
5.0% ±5 digits]
MΩ, 500V
100kΩ
±15 digits
1MΩ
2.0% ±5 digits]
10
2.0% ±5 digits]
100MΩ
2.0% ±5 digits]
1000MΩ
5.0% ±5 digits]
MΩ, 1000V
100kΩ
±15 digits
1MΩ
2.0% ±5 digits]
10
2.0% ±5 digits]
100MΩ
2.0% ±5 digits]
1000MΩ
3.0% ±5 digits]
MΩ, 500V
500KΩ
Source Check
Check on each
Voltage range
0MΩ
Source Check