User Guide
Form number TM61795 Rev 4 July 2011
Page 5
Table 2 Calibration Range Inputs: Model 61-795
Range
Function
Input the calibrator standard value
1
DCV
DC600.0V
2
ACV
AC600.0V
3
190.0Ω
4
0.00Ω
5
LOΩ
19.00Ω
6
LOΩ
0.00Ω
7
LOΩ
5.00Ω
8
LOΩ
0.00Ω
9
250V/MΩ
1000MΩ
10
250V/MΩ
100.0MΩ
11
250V/MΩ
10.00MΩ
12
250V/MΩ
1.000MΩ
13
250V/MΩ
0.000MΩ
14
500V/MΩ
1000MΩ
15
500V/MΩ
100.0MΩ
16
500V/MΩ
10.00MΩ
16
500V/MΩ
1.000MΩ
18
500V/MΩ
0.000MΩ
19
1000V/MΩ
1000MΩ
20
1000V/MΩ
100.0MΩ
21
1000V/MΩ
10.00MΩ
22
1000V/MΩ
1.000MΩ
23
1000V/MΩ
0.000MΩ
Battery Replacement (Refer to Figure 2)
1. Disconnect the test leads from any circuit under test.
2. Press the key to switch off the instrument.
3. Remove the hood and battery compartment cover by using a screwdriver.
4. Remove the batteries replacing them with new ones all of the same type (1.5×6 NEDA 15F IEC R6 JIS
UM-3) (Alkaline batteries are recommended.) making sure of polarity while installing.
5. Install bottom case cover and secure with screws.
6. Install the hood.
Replacing Fuse (Refer to Figure 2)
1. Disconnect the test leads from any circuit under test.
2. Press the key to switch off the instrument.