3. Specifications
Measuring range and accuracy
Function Range Resolution Measuring range Accuracy
Earth 20Ω 0.01Ω 0.00
~ 20.99Ω ±1.5%±0.05Ω
resistance
200Ω 0.1Ω
16.0 ~ 99.9Ω ±2%±0.5Ω
(Auto-ranging) 100.0 ~ 209.9Ω ±3%±2Ω
1Ω
160 ~ 399Ω ±5%±5Ω
1200Ω 400 ~ 599Ω ±10%±10Ω
10Ω 600 ~ 1260Ω
AC current (ACA) 100mA 0.1mA 0.0 ~ 104.9mA ±2%±0.7mA
(sine wave) 1000mA 1mA 80 ~ 1049mA
(50Hz/60Hz) 10A 0.01A 0.80 ~ 10.49mA ±2%
(Auto ranging) 30A 0.1A 8.0 ~ 31.5AmA
* Crest factor <2.5 Accuracy at sine wave +1% (50Hz/60Hz, peak value shall not exceed 60A)
* In the following cases, zero will be displayed on the LCD.
At 20Ω range of Earth resistance function: 0.04Ω or less
* Auto ranging shifts to upper range when the input exceeds 105% of the selected range, and shifts to the lower range when the input
falls under 80% of the lower range.
Operating system Earth resistance function: Constant voltage injection, Current detection,
(Frequency: Approx. 2400Hz)
Dual Integration
AC current function: Successive Approximation (True-RMS)
Display Liquid crystal display with a maximum count of 2099
Over-range indication "OL" is displayed when input exceeds the upper limit of a measuring range
Response time Earth resistance function: Approx. 7 seconds
AC current function: Approx. 2 seconds
Sample rate Approx. once per second
Location for use Altitude 2000m or less, Indoor/ outdoor use
IP protection degree IP40
Temperature & 75°F±5%/Relative humidity 85% or less
humidity range (no condensation)
(guaranteed accuracy)
Operating temperature 14°F ~ 104°F/Relative humidity 85% or less
& humidity range (no condensation)
Storage temperature -4°F ~ 140°F/Relative humidity 85% or less
& humidity range (without batteries, no condensation)
Power source (4) batteries AA
Current consumption Approx. 50mA (max. 100mA)
Measurement time Approx. 12 hours (when manganese battery is used),
Approx. 24 hours (when alkaline battery is used)
Auto power-off Turns power off after 10 minutes of non-use.