User Guide

Page 5
PERFORMANCE VERIFICATIONS
Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning
correctly. If the meter does not conform to any of the listed limits the calibration procedure must be performed.
Performance Verification Preparation
Turn on the calibrator, allow calibrator to warm up. Temperature stabilization should be reached after 30
minutes.
Remove battery cover and use a calibrated meter to ensure the battery measures a minimum of 4.8V DC. If
the battery measures under 4.8V DC, replace the battery before beginning the performance test.
To test the performance of the 61-920 you will need to construct a loop using general resistors with an
accuracy of 0.5% or better of fixed values 0.1, 10, 20 200 1000)
Construct the loop using low
resistive wire. Minimize
connectivity resistance.
Note: Error for the general resistors are not included in the test data of Table 1
If the display reading falls outside of the range shown in the Table 1, steps 1 - 6, the meter does not meet
specification.
Clamp jaws in the center of the test loop.
Table 1: Resistance Performance Verification
Steps
Range Input Low Limit High Limit
1 20
00.10Ω
00.05 00.15
2 20
10.00Ω
9.80 10.20
3 200
20.00Ω
19.1 20.9
4 200
100.0Ω
95.0 105.0
5 1200
200.0Ω
185 215
6 1200
500.0Ω
440 560
Connect the meter to a single wire loop on the calibrator and place the clamp around the center of the wire.
Apply the current listed in Table 2, steps 1 - 3.
If the display reading falls outside of the range shown in the Table 2, steps 1 - 4, the meter does not meet
specification.
Table 2: mA, A Performance Verification
Steps
Range Input Low Limit High Limit
1 100mA 100mA @ 60Hz 97.3 102.7
2 1000mA 1000mA @ 60Hz 980 1020
3 10A 10.0A @ 60Hz 9.80 10.20
4 30A 30.0A @ 60HA 29.4 30.60
Form number TM61920 Rev 4 July 2008