User Manual

36
MCP Process/ISMATEC SA/29.05.07/CB/GP
Analog interface
Pin 1, GND (ground)
Reference potential for all other
inputs
Pin 2, remote
For changing between manual
control and analog interface. For
activating the analog interface,
pin 2 must be connected with pin
1 (GND).
Pin 3, start
- In normal operation (pin 2
open) pin 3 operates the
foot-switch (settings see
page 17)
- In remote operation (pin 2 to
GND) the pump starts when
connected to pin 1 (GND)
Pin 4, direction
In the open position the pump
turns clockwise; when connected
to pin 1 (GND) it turns counter-
clockwise.
Pin 5, speed IN
External speed control
(0 – 5V, 0 – 10V, 0 – 20mA,
4 – 20mA)
Input impedance and input range
can be selected via a dip-switch
inside the pump(see page 38)
Pin 6, input 1
Pin 8, input 2
Digital inputs (TTL-level). Active
when connected to pin 1. Are
only retrievable via RS232 inter-
face; e.g. with a software like the
ProgEdit.
Pin 7, +26V
DC
About +26 V
DC
are available (max.
current 1A).
Interface analogique
Pin 1, GND (masse)
Potentiel de référence pour toutes
les autres entrées
Pin 2, remote
Pour commuter du service manuel
à l’interface analogique.
Pour activer l'interface analo-
gique, le pin 2 doit être connecté
au pin 1 (GND).
Pin 3, start
- en exploitation normale (pin 2
ouvert), le pin 3 sert d’interrupteur
au pied (selection voir en page 17)
- en exploitation à distance (pin
2 sur GND), la pompe se met
en route dès qu’elle est connec-
tée au pin 1 (GND)
Pin 4, direction
Si ouvert, le sens de rotation de
la pompe est celui des aiguilles
d’une montre; si relié avec le
pin 1 (GND), elle tourne dans le
sens contraire des aiguilles d’une
montre.
Pin 5, speed IN
Réglage externe du nombre de
tours (0 – 5V, 0 – 10V, 0 – 20mA,
4 – 20mA) Impédance d‘entrée
et réglage de zone au moyen de
l’interrupteur DIP à l’intérieur de
l’appareil (v.p. 38)
Pin 6, input 1
Pin 8, input 2
Entrées numériques (niveau TTL).
Actifs si reliés au pin 1. Ne peu-
vent être interrogés que par
l’interface RS232, p.e. avec le
logiciel ProgEdit.
Pin 7, +26V
DC
Environ +26 V
DC
sont à dispositi-
on (courant maximal 1A)
Digitale Eingänge (TTL-Pegel)
Digital inputs (TTL-level)
Entrées numériques (niveau TTL)
Pin 2, remote
Pin 3, start
Pin 4, direction
Pin 6, input 1
Pin 8, input 2
Pin 13, speed intern
Analoge Eingänge
Analog inputs
Entrées analogiques
Pin 5, speed IN
0–5 V
DC
/ 0–10 V
DC
0–20 mA / 4–20 mA
Universal Ausgänge (PWM)
Universal outputs (PWM)
Sorties universelles (PWM)
Pin 14, output 1
Pin 15, output 2
Analog-Ausgang
Analog output
Sortie analogique
Pin 9, speed OUT
0-10 V
DC
/ 0-7.2 kHz
Analogschnittstelle
Pin 1, GND (Masse)
Bezugspotential für alle anderen
Eingänge.
Pin 2, remote
Für Umschaltung zwischen ma-
nueller Bedienung und der Ana-
logschnittstelle. Zur Aktivierung
der Analog-Schnittstelle muss
Pin 2 mit Pin 1 (GND) verbunden
werden.
Pin 3, start
- Im Normalbetrieb (Pin 2 offen)
dient Pin 3 als Fußschalter
(Einstellungen siehe Seite 17)
- Im Remote-Betrieb (Pin 2 auf
GND) startet die Pumpe bei Ver-
bindung mit Pin 1 (GND)
Pin 4, direction
Wenn offen, dreht die Pumpe
im Uhrzeigersinn; wenn mit Pin
1 (GND) verbunden, dreht sie im
Gegenuhrzeigersinn
Pin 5, speed IN
Externe Drehzahlsteuerung
(0 – 5V, 0 – 10V, 0 – 20mA,
4 – 20mA)
Eingangsimpedanz und Wahl-
möglichkeiten mittels DIP-Switch
im Geräteinnern (siehe Seite 38)
Pin 6, input 1
Pin 8, input 2
Digitale Eingänge (TTL-Pegel).
Aktiv, wenn mit Pin 1 verbunden.
Können nur über RS232-Schnitt-
stelle abgefragt werden; z.B. mit
einer Software wie ProgEdit .
Pin 7, +26V
DC
Es stehen ca. +26 V
DC
zur Verfü-
gung (max. Strom 1A).
8 7 6 5 4 3 2 1
15 14 13 12 11 10
9
speed IN direction
input 1 start
+26VDC remote
input 2 GND
output 2 speed OUT
output 1 +5VDC
speed intern