User's Manual

IDS Ingegneria Dei Sistemi S.p.A. N°doc: MN/2009/070 - Rev. 1.1
IBIS-L v.02.00 - User manual
37 / 65
All information contained in this document is property of IDS. All rights reserved.
The electric distribution area contains all the system power switches (Fig. 3.22 ). From
left to right we can see:
1. the input fuse from the photovoltaic network;
2. the sensor power supply switch with spy light that illuminates when the sensor is
powered;
3. the linear scanner power supply switch with corresponding spy light;
4. the PC power supply switch;
5. the external power supply source switch (24 V DC) with corresponding spy light;
6. the battery warning light which illuminates when the battery power is low;
7. the battery charge indicator that:
a. shows the charge level of the battery by flashing one of the LEDs when
the battery is being charged or all the switches are off;
b. shows the charge level of the battery by keeping one of the LEDs lit when
the battery is powering one of the instruments but is not being charged;
8. the battery fuse;
9. the switch that controls the power supply input with corresponding spy light and
fuse.
Fig. 3.23 - The power supply module switchboard
The door on the front of the power supply unit provides access to the battery area
containing a pair of 12 V/70 Ah batteries. The back door gives access to the connector
area, which contains the following connections from left to right:
the mains AC input (220V or 110V depending on customer request);
the photovoltaic input, needing an input greater than 24V for battery recharger
with a maximum current of 10A;
the external auxiliary output at 24V;
the 5 pole output to the linear scanner;
the A type USB connection output;
a screw to link the external box (chassis) of the power supply module to the
ground.
8
2
7
5
3
1
6
9
4