To manually send data:
1. When unit is off, press
or .
2. When screen appears, press SET
.
DONE indicates successful data transmission.
Err indicates unsuccessful data transmission.
IMPORTANT To prevent records being erased before
being uploaded to the network, users are reminded to send
data when there are 24 records.
UNIQUE DEVICE ID
Each unit has a unique device ID which can be found on
the device.
TROUBLESHOOTING
This section includes a list of error messages and frequently
asked questions for problems you may encounter with your
blood pressure monitor. If the product is not operating as you
think it should, check here before arranging for servicing.
Symptom
Check This
Remedy
Display
is dim or
will not
light up
Batteries are
exhausted
Replace with new
batteries
Batteries
are inserted
incorrectly
Insert the batteries
correctly
shows
on the
display
Batteries are
low
Replace with new
batteries
Symptom
Check This
Remedy
Flashing
MEM
appears
on the
display
24 or more
records per
user have
not been
uploaded
Upload the data
to the network
(refer to DATA
TRANSMISSION).
Err
appears
on the
display
A measure-
ment error
occurred;
Data
transmission
failed
Relax for a
moment and then
measure again ;
Move closer to the
hub and reattempt
to send data
Err 1,
2, or 3
appears
on the
display
The wrist cuff
is not secure
Refasten the cuff
and then measure
again
Err 4
appears
on the
display
The monitor
detected
motion while
measuring
Movement
can affect the
measurement.
Relax for a
moment and then
measure again
Err 5
appears
on the
display
Pressure is
over
37.3kPa (280
mmHg)
Relax for a
moment and then
measure again
Err 6
appears
on the
display
Deation
period was
too long
Movement
can affect the
measurement.
Relax for a
moment and then
measure again