User's Manual

Page 8
and cleaned by a qualified service technician.
WARNING Liquid can damage electronics inside the monitor. Prevent
liquids from spelling on the monitor.
If liquids are spilled on the monitor:
1. Power down the monitor
2. Disconnect the power plug.
3. Remove battery pack from the monitor.
Note Remove battery pack shall be proceed by trained specialist and
shall not be proceed by user
4. Dry off excess liquid from the monitor.
Note If liquids possibly entered the monitor, remove the monitor from
use until it has been properly dried, inspected, and tested by qualified
service personnel.
5. Reinstall battery pack.
6.
Power on the monitor and verify that the monitor functions normally
before using it.
If liquids enter the USB Box
1. Power down the monitor
2. Disconnect the USB connector.
3. Dry off excess liquid from the USB Box.
Note If liquids possibly entered the USB Box, remove the monitor
from use until it has been properly dried, inspected, and tested by
qualified service personnel.
4. Connect the USB connector
Power on the monitor and verify that the monitor
functions normally
before using it.
WARNING Safety risk. Damaged cords, cables, and accessories can
affect patient and operator safety. Never lift the monitor by the power
supply cord or patient connections. Routinely inspect the AC power cord,
blood pressure cuff, SpO2 cable, and other accessories for strain relief
wear, fraying, or other damage. Replace as necessary.
WARNING Fire and explosion hazard. Do not operate the monitor in the
presence of a flammable anesthetic mixture with air, oxygen, or nitrous
oxide; in oxygen-enriched environments; or in any other potentially
explosive environment.
WARNING The monitor may not function properly if dropped or damaged.
Protect it from severe impact and shock. Do not use the monitor if you
notice any signs of damage. Qualified service personnel must check any
monitor that is dropped or damaged for proper operation before putting the
monitor back into use.
WARNING Defective batteries can damage the monitor. If the battery
shows any sign of damage or cracking, it must be replaced immediately
and only with a battery approved by Fukuda Denshi.
WARNING Electric shock hazard. Do not open the monitor or attempt
repairs. The monitor has no user-serviceable internal parts. Only perform
routine cleaning and maintenance procedures specifically described in this
manual. Inspection and servicing of internal parts shall only be performed
by qualified service personnel.