WARNING Inaccurate measurement risk. Do not expose to temperatures
higher than 42.9°C
WARNING Inaccurate measurement risk. Do not use the monitor on
patients who are on heart-lung machines.
WARNING Use the monitor only as described in this directions for use. Do
not use the monitor on patients as described in the Contraindications.
WARNING Inaccurate measurement risk. Do not use the monitor on
patients who are experiencing convulsions or tremors.
WARNING Wall mounted equipment and accessories must be installed in
accordance with accompanying instructions. Fukuda Denshi is not
responsible for the integrity of any installation not performed by authorized
Fukuda Denshi service personnel. Contact an authorized Fukuda Denshi
service representative or other qualified service personnel to ensure
professional installation for safety and reliability of any mounting
accessory.
WARNING Do not place the monitor in any position that might cause it to
fall on the patient.
WARNING Fukuda Denshi is not responsible for the integrity of a facilitys
power. If the integrity of a facilitys power or protective earth conductor is in
doubt, always operate the monitor on battery power alone when it is
attached to a patient.
WARNING Equipment damage and personal injury risk. When
transporting the monitor on a mobile stand, properly secure all patient
cables and cords to keep them clear of the wheels and to minimize trip
hazards.
WARNING For operator and patient safety, peripheral equipment and
accessories that can come in direct patient contact must comply with all
applicable safety, EMC, and regulatory requirements.
WARNING All signal input and output (I/O) connectors are intended for
connection of only devices complying with IEC 60601-1, or other IEC
standards (for example, IEC 60950), as applicable to the monitor.
Connecting additional devices to the monitor may increase chassis or
patient leakage currents. To maintain operator and patient safety, consider
the requirements of IEC 60601-1. Measure the leakage currents to confirm
that no electric shock hazard exists.
WARNING Equipment failure and patient harm risk. Do not cover the air
intake or exhaust vents on the rear and base of the monitor. Covering
these vents could cause overheating of the monitor or muffing of alarms.
WARNING This equipment is not suitable for use in the presence of
electro-surgery.
WARNING Cross-contamination or nosocomial infection risk. Clean and
maintenance the monitor on a routine basis according to your facilitys
protocol and standards or local regulations. Thorough hand-washing
before and after contact with patients greatly reduce the risk of
cross-contamination and nosocomial infection.
WARNING Do not modify this equipment without authorization of the
manufacturer