User's Manual

Page 11
settings.
Caution Never move the monitor or mobile stand by pullimg on any of the
cords as this may cause the monitor to tip over or may damage the cord.
Never pull on the power cord when removing it from the power outlet.
When disconnecting the power cord, always grasp the attachment plug
and not the cord. Keep the cord away from liquids, heat, and sharp edges.
Replace the power cord if the strain relief or cord insulation is damaged or
being to separate from the attachment plug.
Caution Use only the Fukuda Denshi USB client cable to connect a USB
Box to the USB port.
Caution If the touchscreen is not responding properly, refer to the
troubleshooting section. If the problem cannot be resolved, discontinue
use of the monitor and contact an authorized Fukuda Denshi service
center or qualified service personnel.
Caution Summation of leakage currents when several items of ME
EQUIPMENT are interconnected
Caution The electrocardiograph incorporates a means to protect the
patient against burns when used with HIGH FREQUENCY(HF)
SURGICAL EQUIPMENT
Caution Cleaning and maintenance procedures
l When cleaning the touch panel, never use strong-acidic cleaning
solution.
l A special coating is applied to the surface of the touch panel. Do not
wipe the surface with a cloth or gauze with coarse texture. Wipe the
surface with an eyeglass cleaning cloth.
l Clean the equipment frequently so stains can be removed easily.
l To prevent injury, it is recommended to wear gloves when cleaning
the equipment.
l Do not allow liquids or cleaning solution to enter the equipment or
connectors.
l Do not use organic solvents, thinner, toluene and benzene to avoid
damaging the resin case.
l Do not polish the housing with abrasive or chemical cleaner.
l When sterilizing the entire room using a spray solution, pay close
attention not to have liquids get into the equipment or connectors,
The surface resin coating may be damaged, resulting in
discoloration, scratches, or other problems.
Example: Do not use chemical cloth, scrub brush, abrasive, polishing
powder, hot water, volatile solvent and chemicals (cleanser, thinner,
benzine, benzol, and synthetic detergent for house and furniture), or
sharp-edged tools.
l Do not open the housing.
l Avoid alcohol or other liquids from getting into the equipment.
l For safe operation of the equipment, regular inspection and
maintenance are required. Once a year, check all cables, devices,
and accessories for damage, earth impedance, earth and leakage
currents, and all alarm functions. Also, ensure that all safety labels
are legible. Maintain a record of these safety inspections.