Monitoring
Main Screen
Patient information, battery life, dates and countdown timer are displayed at the top
of the page in the Monitor Mode. The countdown timer in green indicates the
remaining time for next auto blood pressure check. If the number is absent, that
means the auto check function is disabled.
Other displayed values are described below:
l HR: heart rate. A heart icon and a beep will be generated every time the ECG peak
is detected. The numbers under bpm on the right indicate the normal range of heart
rate. If the heart rate exceeds the normal range, the warning sound and icon will be
triggered.
l PR: pulse rate (displayed when SpO2 is selected without ECG channel). A beep will
be generated every time the SpO2 peak is detected.
l SpO2: oxygen saturation
¡ The five blocks on the left show the level of heart beat strength. The three
blocks on the top show the level of the signal strength.
¡ The number beside the percentage sign (%) indicates the normal range. If the
oxygen saturation exceeds the normal range, the warning sound and icon will
be triggered.
l NIBP: blood pressure
¡ The five blocks on the left show the strength level of the blood flow
¡ Time: the time at which the checkup is completed
¡ Sys in orange: Systolic pressure. The small number on the side indicates the
normal range of systolic pressure. If the systolic pressure exceeds the normal
range, the warning sound and icon will be triggered.
¡ Dia in orange: Diastolic pressure. The small number on the side indicates the
normal range of diastolic pressure. If the diastolic pressure exceeds the normal
range, the warning sound and icon will be triggered.
¡ CUFF: the pressure of the cuff
¡ Press START/STOP to start the checkup. After start, press START/STOP
again can stop the checkup.
¡ Press any place other than the START/STOP button to enter the page with
detail information of the NIBP check.