User's Manual
Page 49
Alarms
The monitor presents physiological alarms and technical alarms. Physiological alarms occur
when vital sign measurements fall outside of set alarm limits.
Alarm types
Type Priority Color Alarm audio tone
Heart Rate, Respiration rate, SpO2, NIBP, or
Pulse rate limit exceeded
High Red 5-pulse tone
Some technical alarms Low Blue 1-pulse tone
Non-latching alarm
The alarm of this monitor is non-latch type.
The alarm stops when the vital sign returns within the set alarm limits.
However, alarms that occurred in the past can be confirmed in the alarm history screen.
If there is an unconfirmed alarm, change the Alarm History button to red and inform it
Alarm notification locations
WARNING If you are relying on visual alarm notifications, maintain a clear
line of sight with the monitor. If you are relying on audio alarm notifications,
ensure that you can hear audio alarms from where you are. Set the
volume as needed considering the environment and ambient noise levels.
LED light bar
The light bar on the top of the monitor illuminates as follows:
l Flashing red for high priority alarms
l Constant blue for low priority alarms
Main screen