User's Manual
Page 50
Main screen notifications
Notification Description
Information area
Audio pause button appears when alarm occurs. If the audio
pause button is pressed, a timer countdown appears.
If monitoring is started within 1 minute (or 2 minutes), or alarm
pause button is pressed, alarm does not work. During that time
countdown timer is displayed. (Depending on the setting in 1 or 2
minutes)
Parameter area
Display alarm information with red (high priority) or blue (low
priority) frame.
If multiple alarms are active, the most recent alarm message
appears.
You can confirmation each alarm on the alarm history screen.
Menu area
Alarm History button changes color to red. By pressing this button
you can confirmation alarm on the alarm history screen, and the
button changes color to black.
Icons on the Main screen
Icons in the parameter area
Icon Name and status
Alarm off.
No visual or audio alarms occur for this limit.
WARNING Alarm does not work during 12-lead ECG measurement. Also,
the countdown timer of Alarm Pause is displayed from the start of
monitoring and returning to monitoring from the 12-lead ECG
measurement during which the alarm will not operate
Icons in the information area
Icon Name and status