alarm.
If you turn off alarm, no visual or audio alarm signals will occur for those parameters.
If the alarm is off, the icon on the Main screen in the parameter area.
Default alarm limit
You can set the alarm limit to default with touch the Default button.
Default alarm limits are follows.
Parameter Unit Lower Limit Upper Limit
HR/PR
BPM 40 120
SpO2
% 90 100
Resp
/min 5 30
NIBP Sys
mmHg 80 180
NIBP Dia
mmHg 40 100
NOTE Whether the alarm limit at patient admit to the default value or to use the previous
patient value depends on System setting
Modify audio alarm notification
You can modify the volume of all audio alarm.
WARNING The alarm volume should be loud enough for you to hear it
from where you are. Set the volume considering the environment and
ambient noise level.
l Touch the Menu key.
l Touch the Sound/List key.
l On the Sound/List screen, modify alarm volume level.