ECG Monitor Setting (Lead, Sensitivity, Filter, Pacing Detect)
1. Touch the Menu key.
2. Touch the ECG Monitor key.
3. You can configure the lead, sensitivity, filter, and pacing detect.
4. Select ECG Lead.
l The ECG function can use 3 wire and 10 wire lead.
l Only I, II, III leads can be selected regardless of which lead is used.
l Choose the lead with the QRS of ECG, the largest on the monitor.
5. Select Sensitivity
The height of the vertical ruler that appears to the left of the ECG waveform indicates 1
mV amplitude and is 10 mm high if x1 is chosen.
6. Select the Filter in accordance with the intended use.
l Diag.: Use when observing the ST change due to ischemia.
l Monitor: Usually, use this mode.
l Drift Cut: Select when using in an environment with a lot of ac noise. Waveform
distortion increases.
Note This DS-101 does not have arrhythmia detection. Please
carefully observe the patient's condition by using upper / lower limit
alarm of Heart Rate.