Note Detection of ischemia is the interpretation of the clinician only,
the DS-101 does not provide automated ischemia detection.
Note It is normal for the ECG baseline to wander slightly in Diag. Filter.
7. Turn on the Hum Filter with the intended use.
Turn on when AC noises enter even if electrode and lead wires are adjusted.
8. Turn on the Pacing Detect with the intended use.
l Used for pacemaker patients.
l Pacing Detection is on, the DS-101 displays pacemaker signals exactly as they are
captured.
WARNING PACEMAKER PATIENTS. See this manual for disclosure of
the pacemaker pulse rejection capability of this instrument”
WARNING Heart Rate may continue to count the pacemaker rate during
occurrences of cardiac arrest or some arrhythmias. Do not rely entirely
upon rate meter alarms. Keep pacemaker patients under close
surveillance
RESP (Respiration)
RESP frame
From the RESP frame, you can measure Respiration Rate (RR).
Located in the center of the Main screen, the RESP frame contains data and features
relevant to Respiration.
Main Screen
Large Screen
l You can check the Respiration Rate
l You can see the synchronized chest motion with RESP wave (Main Screen) or 5 level
bar (Large Screen).
l You can see Respiration Rate alarm limits. (Upper and Lower), right of the RR.
l If alarm is off, you can see alarm off mark.
l When either of If RR goes out of the alarm range, the red LED flashes, an alarm tone
sounds and an alarm message is displayed.