User's Manual
Page 63
Pulse wave display
l The Pulse waveform is always display on the bottom of the Main screen, expect of ECG
1ch, ECG cascade modes.
ECG & SpO2 mode
SpO2 mode
l Amplitude of the pulse wave is automatically adjusted.
l Touch the screen to fix the waveform.
PR display
When pulse only mode is selected, Heart Rate (HR) display changes to Pulse Rate (PR)
display. You can check the Pulse Rate on ECG frame.
HR display PR display
Note When the pulse only mode is selected, ECG and RESP related alarm
are turned off. (Including ECG and RESP lead off warning)
SpO2 and PR Alarm
1. Touch the Menu key.
2. Touch the Alarm Setting key.
3. SpO2 % is displayed in the SpO2 key and the green line on the bar graph.
4. To set the SpO2 alarm limits, use the arrow key on the right side of the screen to follow
these steps:
l Touch the Lower Limit key, change limit using arrow key.
l Touch the Upper Limit key, change limit using allow key.
l Alarm limit is changed immediately.
5. In the pulse only mode, use the arrow on the right side of the screen to set the alarm
limits of the Pulse Rate
6. If you want to turn off the SpO2 alarm, press the SpO2 key to make it black.
7. If you want to turn off the PR alarm, press the HR/PR key to make it black.