User's Manual

Page 85
Respiration
Method Impedance Method
Frequency Characteristic 1.5Hz
Current 100μA and below (at 64kHz ±5%)
Measurement Range 0, 4 to 150bpm
Measurement Accuracy ±3rpm
Waveform Size Selection 1/5, 1/2, 1, 2, 4 (1 means 10mm/1Ω)
Seep Speed 1.6mm/sec
SpO2 (Arterial Oxygen Saturation)
Measurement Method 2 Wavelength Pulse Wave Method
Wavelength: Approx. 660nm (red light), Approx.
940nm (Infrared Light)
Output: 15mW and below
Measurement Range 40 to 100%
Resolution 1%
Measurement Accuracy Adult: ±3% when 70 to 100%
PR Measurement Range 20 to 250bpm
PR Accuracy ±3bpm
Measurement Response Time 6 to 7 sec
Measurement Value Update Time 1 sec
Pulse Waveform Size Selection Automatic
Pulse Waveform Seep Speed 25mm/sec
NIBP (Non-Invasive Blood Pressure)
Measurement Method Oscillometric Method
Measurement Range 30 to 280mmHg/4 to 37.3kPa
Resolution 1mmHg
Static Pressure Accuracy ±3mmHg/0.4kPa
BP Measurement Error according to the Clinical Performance Test
Mean Error Within ±5mmHg
Standard Deviation of Error 8mmHg or below
Error of Cuff Pressure Display Within ±3mmHg
PR Measurement Range 40 to 240bpm
PR Accuracy ±2% or ±2bpm (whichever greater)
Safety Mechanism 300mmHg or above