User's Manual
Page 93
Check if the cuff is properly attached to the patient, or cuff size is correct.
Cause 3
The air hose is disconnected.
Solution
Check if the air hose is tightly connected, and then measure again. If the same message is displayed
again, air may be internally leaked.
Contact our service representative.
Temperature
No value display on the screen
Cause 1
Data is not transferring successfully
Solution
Please contact with PC directly with close enough distance and place in the correct location
Cause 2
Thermometer value has been read once
Solution
Please measure and read the value again.