Users Manual

iGlucose® 17
auto code
turning the meter on / off
The iGlucose Blood Glucose Meter will automatically detect the
code number on the strip. You do not have to manually enter a code
number.
To power on the iGlucose Meter, press and hold the “Home”
button until the meter turns on. The display will light up. In the
next few seconds the meter will display “Searching” while the
date and time automatically synchronize. When done, a message
“Date & Time Synchronized” will appear and the meter will enter
“Standby Mode.”
When you open a new vial of iGlucose® Blood Glucose Test Strips,
record the date on the vial. Discard the vial of test strips after 3
months from opening.
Store the iGlucose Test Strips, between 39-8F (or 4-30°C) and
in a location 10-90% relative humidity. Do not expose to direct
sunlight or heat.
Storage of strips near bleach as well as bleach containing
products will aect the results of the iGlucose Test Strips.
Do not perform testing immediately after moving from one
ambient temperature to another (e.g. after coming indoors from
the outside). Allow 45 minutes for the temperature of the meter
and the test strips to equilibrate. The need to wait 45 minutes is
required when for example, a meter is kept for a while in a car on a
hot afternoon and then brought into an air conditioned room for
testing or when a meter is kept outside on a snowy day and then
brought into a warm room.
For more information, please refer to the iGlucose Test Strips
Package Insert.
The use of other test strips may cause strip error.