Immunity test IEC 60601 test level Compliance
level
Electromagnetic environment-guidance
Radiated RF
IEC 61000-4-3
3 Vrms
80MHz to 2.5GHz
3V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the SYSTEM, including cables, than
the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
P
V
d ]
5.3
[
1
=
P
E
d ]
5.3
[
1
=
80MHz to 800MHz
P
E
d ]
7
[
1
=
800MHz to 2.5GHz
Where P is the maximum output power rating
of the transmitter in watts(W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in
each frequency range.
Interference may occur in the vicinity of
equipment marked wigh the following
symbol:
Recommended separation distances between portable and mobile RF communications equipment and the
ME EQUIPMENT and ME SYSTEM – for ME EQUIPMENT and ME SYSTEM that are not
LIFE-SUPPORTING.
Recommended separation distances between portable and mobile RF communications equipment and the
HS5
The HS5 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the HS5 can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters ) and the HS5
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150kHz to 80 MHz 80MHz to 800MHz 800MHz to 2.5GHz