Operation Manual
PROBLEMA POSSIBILE CAUSA SOLUZIONE
Battery is less than 25% Charge the batteryLow Battery
Blood pressure is outside of measurement range Retest, make sure your blood pressure is within
measurement range
Wrist or monitor was moved during test Retest, make sure not to move your wrist or the monitor
The cuff does not inflate properly or pressure
falls quickly during test
Review the cuff application instructions and retest
Irregular heartbeat (arrhythmia) It is inappropriate for people with serious arrhythmia to
use this monitor .
Check with physician.
The cuff was not properly applied
The cuff position was not correct or it was not
properly tightened.
Review the cuff application instructions and retest.
Review body posture instructions and retest
Retest when calm; avoid speaking or movement during
the test
Body posture was not correct during testing
Speaking, moving wrist or body, being angry,
excited or nervous during test
Review the cuff application instructions and retest
Display reads
“ERROR”
Display reads
an abnormal
result
Bluetooth
connection
unstable
Bluetooth connection unsuccessful, monitor is
abnormal, or strong electromagnetic interference
is present
Reset iOS device. Reset monitor by pressing the
START/STOP button about 10s. Make sure the monitor
and iOS device are away from other electrical equipment.
Please see GENERAL SAFETY AND PRECAUTIONS.
Incorrect operation or strong electromagnetic
interference
Retest, make sure not to move your wrist or the monitor
No response
TROUBLESHOOTING
12