Guidance and manufacturer’s declaration - electromagnetic immunity
The Wireless Blood Pressure Wrist Monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the Wireless Blood Pressure Wrist Monitor should assure that it is used in such an environment.
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
IMMUNITY test
IEC 60601test level
Compliance level Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 Vrms 150 kHz to
80 MHz
3 V
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the monitor,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter .
Recommended separation distance:
Radiated RF
IEC 61000-4-3
3 V/m 80 MHz to
2.5 GHz
Pd 2.1=
Pd 2.1=
80 MHz to 800 MHz
Pd 3.2=
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey,a should be less than the
compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
Table 3