Operation Manual

For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Pd 2.1= Pd 2.1= Pd 3.2=
Recommended separation distances between portable and mobile RF communications equipment and the Wireless Blood Pressure
Wrist Monitor
The Wireless Blood Pressure Wrist Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Wireless Blood Pressure Wrist Monitor can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the monitor as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.12 0.230.120.01
0.38 0.730.380.1
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
1.2 2.31.21
3.8 7.33.810
12 2312100
Table 4
22