iHealth® PT3SBT Wireless No-Touch Forehead Thermometer FCC ID:SLRPT3SBT IC:10913A-PT3SBT 1
iHealth® Wireless No-Touch Forehead Thermometer (PT3SBT) OWNER’S MANUAL Table of content INTRODUCTION ................................................................................................................................. 3 Indication for use .............................................................................................................................. 3 Safety Precautions...............................................................................................................
INTRODUCTION Thank you for choosing our product. This product is a high-tech infra-red (IR) thermometer designed to take human body temperature by measuring the energy of IR emitted from the forehead and it can transmit the temperature to a smart device with Bluetooth. The product will help you to assess you and your family members’ health conditions easily and quickly.
such tissues will affect temperature conduction of the body. 7. If there is a temperature difference between the thermometer storage area and the new ambient environment around the subject, please let the thermometer sit within the new environment for 30 minutes before taking the measurement. 8. Do not measure body temperature immediately after consuming a drug that raises body temperature. Temperature readings taken at this time will not be accurate. 9.
3. Do not directly touch the probe with your fingers or blow on it. Measurements taken using a damaged or dirty IR probe may be inaccurate. 4. Keep the product at a place inaccessible to children to prevent children from swallowing the batteries or small parts. 5. Do not throw the thermometer or batteries into fire to prevent explosions. 6. Remove the batteries from the thermometer if the device will not be used for more than one month. 7. If you are allergic to plastic/rubber, please don’t use this device.
designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation.
LED screen instructions Device dimensions: 5.39 in x 1.34 in x 1.54 in Screen dimensions: 1.01 in x 0.75 in Product weight: 90 g Product performance 1. Measurement position: Centre of the forehead surface 2. Measurement distance: ≤1.18 in(3 cm) size AAA batteries 3. Power source: DC 3V: 2 x 1.5V 4. Measurement range: 89.6℉-109.2℉(32℃-42.9℃) 5. Measurement precision: ±0.4℉(±0.2℃) within 93.2℉-107.6℉(34℃ -42℃), and±0.5℉(±0.3℃) for other temperature ranges. 6. Resolution: 0.1℉(0.1℃) 7.
ASTM laboratory accuracy requirements in the display range of 37 to 39°C (98 to 102°F) for IR thermometers is ± 0.2 °C ( ± 0.4°F), whereas for mercury-in-glass and electronic thermometers, the requirement per ASTM Standards E667-86 and E1112-86 is ±0.1 °C (±0.2°F). Expected use and scope of applications This product mainly uses infrared sensors to measure the temperature of the human forehead and transmit the temperature to a smart device with Bluetooth. It can be used for babies, children, and adults.
② Gently press the measurement button to start the measurement. ③ The device will vibrate once a reading is obtained. Once the measurement is successful, the reading will be displayed on the screen. The whole time is about 1s. If the measurement is failed, the device will not vibrate the screen but display instead. If the [ ] icon on the thermometer screen is always on, the measured temperature data can be uploaded to the smart device. Possible reasons for measurement errors include: A.
(4) Unit switching 1. When the device is powered OFF, press and hold the measurement button for 8 seconds to enter unit switching mode. Under this mode, both ℃ and ℉ should flash at the same time. 2. Under this mode, press the measurement button to switch to Fahrenheit or Celsius. The corresponding symbol of the selected unit after switching will flash. 3. After selecting the unit, press and hold the measurement button for 8 seconds to leave the unit switching mode.
SYNC TIME AND RESULTS VIA BLUETOOTH Prior to first use, scan the QR code to install the iHealth App. Note: The App works with both iOS and Android device. For a complete list of compatible devices, visit our support on page on www.ihealthlabs.eu. Connect to iOS Device via Bluetooth 1. Launch the iHealth App from your iOS device. 2. Enable Bluetooth on your iOS device: Slide gently upward the iPhone screen from to open the Control Center, and click on the . 3.
NOTE: During the upgrade process, the Bluetooth icon on the thermometer is always on and the three horizontal bars are displayed in cycles. Product errors and troubleshooting Problem Item to detect The application cannot find the product. Solution Battery depleted Replace old batteries with new ones Batteries have been installed with the wrong polarity Batteries are not installed properly Take out the batteries and re-install them correctly.
healthy forehead temperature to correctly determine the temperature. Body temperature runs approximately from 95.9°F to100°F(35.5℃-37.8℃). To determine if one has a fever, compare the temperature detected with the person's normal temperature. A rise over the reference body temperature of 1.8℉(1℃) or more is generally an indication of fever. Different measurement sites (rectal, axillary, oral, frontal, auricular) will give different readings.
Maintenance 1. This company has not authorized any agency or individual to carry out product repairs or maintenance. Do not attempt to disassemble or modify the thermometer if you suspect functional issues with the device. 2. The IR thermometer is an extremely precise instrument. Any improper maintenance, disassembly, or modification may lead to inaccuracies of the product measurements. 3. If you suspect any product issues during the warranty period, please contact customer services for subsequent handling.
1 IR Thermometer 1 Instruction manual 2 AAA batteries 1 Quick user guide 1 Storage Box Standard icons Symbol for “THE OPERATION GUIDE MUST BE READ” The sign background color: blue. The sign graphical symbol: white. The batteries and electronic instruments must be disposed of in accordance with the locally applicable regulation, not with domestic waste.
(Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance), IEC 60601-1-2:2014(Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance Collateral standard: Electromagnetic compatibility - Requirements and Tests), ISO 80601-2-56:2017/AC:2018( Medical Electrical Equipment -- Part 2-56: Particular Requirements For The Basic Safety And Essential Performance Of clinical thermometers for body temperature measurement),
Table 3 – Proximity fields from RF wireless communications equipment Test frequency (MHz) 385 450 710 745 780 810 870 930 1720 1845 1970 2450 5240 5500 5785 Band (MHz) 380-390 430-470 704-787 Immunity test levels Professional healthcare facility environment Pulse modulation 18Hz, 27V/m FM, ±5kHz deviation, 1kHz sine, 28V/m Pulse modulation 217Hz, 9V/m 800-960 Pulse modulation 18Hz, 28V/m 1700-1990 Pulse modulation 217Hz, 28V/m 2400-2570 5100-5800 Pulse modulation 217Hz, 28V/m Pulse modulation 217Hz,
www.ihealthlabs.com EC REP iHealthLabs Europe SAR 36 Rue de Ponthieu, 75008, Paris, France support@ihealthlabs.eu www.ihealthlabs.eu ANDON HEALTH CO., LTD. No. 3 Jinping Street, Ya An Road, Nankai District, Tianjin 300190, China Tel: 86-22-60526161 This product can be used across EU member states. EU Regulatory Conformance Hereby, iHealth Labs Inc. declares that this device is in compliance with the essential requirements and other relevant provisions of Directive 2014/53/EU.