User guide
39
Display unit: this sets the display units for the
calorimetric results. Select them from the list.
The selected range of units applies to the display
of energy values [J], calorifi c values [J/g] and ther-
mal capacities [J/K].
Reference calorifi c value: Input here the ca-
lorifi c value of the substance used for calibration.
As a rule this substance is benzoic acid with a
calorifi c value certifi ed by the manufacturer.
„Measurement“ tab
Double determination: If this menu item is
activated, two measurements are basically com-
bined into a group.
The suffi x ЖRP_x is added onto the end of the
testing code. (X = 0, 1 or 2).
In the process, the fi rst two calorifi c values are
checked to ensure they fall within 120J. If the
difference between them is greater than 120J,
a warning appears that a third measurement is
required.
The third measurement must then fall within
120J of one of the fi rst two measurements.
If the fi rst two measurements fall within 120J of
each other, the next measurement creates a new
group.
Measurement
Other Settings
Select from the list:
The selected standard will be applied to all the
subsequent measurements.
Evaluated measurements will comply with this
standard: if a measurement has already been
evaluated, the evaluation procedure will be used
for any subsequent re-evaluation. Otherwise the
current setting for the evaluation procedure will
be used.
Measurements evaluated previously are not af-
fected by this change.