• The equipment may heat up by stirring.
• Gradually increase the speed.
Before you fill the reactor vessel, ensure
that the reagents used do not corrode
the FTFE and FKM seal.
• Ensure that the external temperature sensor is inserted in
the media to a depth of at least 20 mm.
• The external temperature sensor must always be inserted
in the media when connected.
• Only use IKA
®
approved accesories!
• Use only original IKA
®
spare parts!
Risk of burns! Exercise caution when
touching the heating block and the glass
vessel. The heating block can reach temperatures in excess
of 230 °C and the glass vessel can reach 120 °C. The cover
of the vessel could reach high temperature when you boil
liquid for long time. Pay attention to the residual heat af-
ter switching off.
• Ensure that the thermostat used for tempering is fully ef-
ficient. A defective thermostat can cause uncontrolledre-
actions.
Only process and heat up any media that
has a flash point higher than the adjust-
ed safe temperature limit that has been set (50...210 ºC).
The safe temperature limit must always be set to at least
25 °C lower than the fire point of the media used.
Use
The IKA
®
LR 1000 Control system is a compact reactor
system. It has been designed to simulate and optimize
chemical reaction processes and can also be used for
mixing, dispersion and homogenization in modeling pro-
cesses.
Operating mode: Tabletop device.
Range of use (indoor use only)
- Laboratories - Schools
- Pharmacies - Universities
Correct use
This equipment is suitable for use in all areas except:
- Residential areas
-
Areas that are connected directly to a low-voltage supply
network that also supplies residential areas.
The safety of the user cannot be guaranteed:
- if the instrument is operated with accessories that are not
supplied or recommended by the manufacturer
- if the instrument is operated improperly or contrary to the
manufacturer’s specifications
- if the instrument or the printed circuit board are modified
by third parties.
DANGER
DANGER
WARNING
• Unpacking
- Please unpack the device carefully
- In the case of any damage a detailed report must be
sent immediately (post, rail or forwarder).
• Delivery scope
- LR 1000 control base
- LR 1000 reactor vessel
- LR 2000.61 sensor receptacle
- 2 hose connecting pieces
- Temperature sensor PT 100.5
- USB cable
- Mains cable
Unpacking
- Tools kit see Fig. 1
- Operating instructions
- Warranty card
- Decontamination clearance certificate.
Fig. 1