User Manual

35
Accessories
Disperser T 18 digital
Disperser T 25 digital
Disperser adaptor (M36x1.5)
O-ring (Ø 18 x 3.55, FKM)
O-ring (Ø 11.6 x 3, FKM)
Warranty
In accordance with IKA warranty conditions, the warranty
period is 24 months. For claims under the warranty please
contact your local dealer. You may also send the machine
direct to our factory, enclosing the delivery invoice and giv-
ing reasons for the claim. You will be liable for freight costs.
Materials in contact with medium/laboratory reactor
Reactor cover Stainless steel 1.4571 / RCI 316 L
Reactor cover bottom Stainless steel 1.4571 / RCI 316 L
Reactor vessel Borosilicate glass 3.3
Anchor stirrer Stainless steel 1.4571 / RCI 316 L
Scraper PTFE / Stainless steel 1.4571 / RCI 316 L
O-ring FKM
Shaft seal PTFE
O-ring (Ø 41 x 1.9, FKM)
O-ring (Ø 26 x 2, FKM)
Adaptor for PH meter
Anchor Stirrer with PTFE scraper
The warranty does not cover worn out parts, nor does it
apply to faults resulting from improper use, insucient care
or maintenance not carried out in accordance with the in-
structions in this operating manual.
Error 17 • SAFE Temp is set at a temperature
lower than the actual heating block
temperature
Heating off • Allow the unit to cool
• Set the SAFE Temp at a higher temperature
Error 26 • Fault in the internal or safety
temperature sensors or measurement
circuit
Heating off • Substitude the internal and safety temperature
sensor.
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.