MEDRAD
®
Salient Operation Manual 17
3.3.6. Remote Control Symbols and Buttons
3.3.7. Injector Visual Indicators
Event LEDs at rear of injector head LCD Display
Injector Error Condition
1
Flash Red Red Status Bar
Warning Flash Orange Orange Status Bar
ARMED State Flash Green Flashing ARM symbol
INJECTING State Rotate Green Flashing Syringe
INJECT COMPLETE State Stop rotating (orange if adaptive
flow occurred, green otherwise)
Nothing
Touch screen Button press Nothing Nothing
Motor movement Rotate green Animated Syringe
Adaptive Flow
(Pressure Limiting)
Rotate orange Orange Status Bar
1
Notes: There are no alarm presets in the injector.
All error conditions indicate a fault with the injector and the injection has stopped. Warnings and other
indicators indicate continued but changed operation of the injector.
Battery Low LED
- If the
battery is low this light will
shine continuously and the
remote will emit a beep. If this
occurs the batteries should be
replaced. See Section 6.2.2.
The light may also come on
when the red or green button is
pressed, but this is not an
indication of a low battery.
Pause button
- When the
injector is injecting, pressing this
button stops the current injection
immediately. The button has no
effect if the injector is not
injecting.
Transmitting LED - When
lit, this LED indicates that
the remote is transmitting
to the injector.
Start buttonStarts or
continues an injection. The
button has no effect if the
injector is not armed