Operation Manual
Table Of Contents
- 1. Quick Start
- 2. Introduction
- 3. Getting Started
- 3.1. Installation
- 3.2. Injector Features
- 3.3. Symbols
- 3.3.1. Symbols used in this manual
- 3.3.2. Symbols used on labelling
- 3.3.3. Handswitch and Footswitch
- 3.3.4. Icons and Buttons used on the Injector Touch Screen Graphical Interface
- 3.3.5. Injector Buttons
- 3.3.6. Remote Control Symbols and Buttons
- 3.3.7. Injector Visual Indicators
- 3.3.8. Icons and Buttons used on the RCU Touch Screen Graphical Interface
- 4. Clinical Information
- 5. Operation Description
- 5.1. Turning the System On for the First Time
- 5.2. Syringe, Quick Fill Tube, Fill Spike, and Connector Tubes
- 5.3. Filling a Syringe Using a Quick Fill Tube (QFT) or Fill Spike on the Injector
- 5.4. [Optional] Off-line Filling
- 5.5. Connector Tube Installation
- 5.6. Dismounting a Syringe
- 5.7. Performing an injection
- 5.8. Programming
- 5.9. Options Menu
- 5.10. [RCU models] Operation Description with Remote Control Unit
- 5.10.1. RCU Features–
- 5.10.2. The Injector and the RCU
- 5.10.3. Turning the System On / Off
- 5.10.4. Injector Controller Application
- 5.10.5. Filling State
- 5.10.6. Routine Protocol
- 5.10.7. Arming
- 5.10.8. Check for Air
- 5.10.9. Starting the Injection
- 5.10.10. Injection in Progress
- 5.10.11. Injection Control Buttons
- 5.10.12. Injection Review
- 5.10.13. Protocol Edit
- 5.10.14. Edit Lock
- 5.10.15. Parameter Value Edit
- 5.10.16. Injecting remaining volume
- 5.10.17. Edit Phase Type
- 5.10.18. Open Protocol
- 5.10.19. Protocol Save
- 5.10.20. Common Injecting Alarms / Warnings
- 5.10.21. RCU Manager
- 5.10.22. RCU Setup
- 5.10.23. History Viewer
- 5.10.24. Protocol Manager
- 5.10.25. eGFR (glomerular filtration rate) Calculator
- 5.11. [Optional] DC200 Scanner Interface
- 5.12. [Optional] Protocol Assistance Tool (PAT)
- 6. Cleaning and Maintenance
- 7. Trouble-shooting
- 8. Specifications
- 8.1. Dimensions
- 8.2. Mechanical
- 8.3. Functional
- 8.4. Controls
- 8.5. Environmental
- 8.6. Electrical
- 8.7. Connectors
- 8.8. IEC 60601-1 Classifications
- 8.9. Ground Continuity
- 8.10. RoHS Statement
- 8.11. IEC60601-1-2:2014 (4th Ed) Compliance
- 8.12. Models
- 8.13. Accessories
- 8.14. Manufacturer’s Software License Information
- 8.15. FCC Statement
- 9. Index
MEDRAD
®
Salient Operation Manual 25
4.4. Warming the Contrast: Heat Maintainer
For improved patient comfort and lower
viscosity, contrast manufacturers recommend
warming contrast to body temperature before
use (i.e. 37°C). It is also recommended that the
contrast be warmed for at least 2 hours prior to
use, because heat permeates contrast very
slowly.
Imaxeon provides a syringe heat maintainer
part number DC022, which can be used to
maintain the heat of the fluid in the syringe.
The warmer is powered from a port located on
the underside of the injector head unit.
[Dual Only] The Dual injector includes two
connectors for saline and contrast heat
maintainers.
WARNING
Use only the specified Imaxeon
heat maintainer (DC022) with the
Salient injector. Do not connect
any other device to the injector
heat maintainer connector.
WARNING
The syringe heat maintainer
incorporates a red failure indicator
light; if indicator light is lit while
on the syringe, remove the heat
maintainer. This indicates that the
syringe heat maintainer is faulty
and the built-in safety circuitry
has disabled the heating
mechanism.
Heat Maintainer Connector