Operation Manual

Table Of Contents
MEDRAD
®
Salient Operation Manual 31
5.3.1. Manual and Auto Filling Procedure
Manual filling Integral AutoFill with a QFT
1. Mount a new syringe on the injector
2.
Select on the touch screen
3. The plunger will automatically purge the air from the syringe.
4. Push the QFT onto the end of the syringe. Do not install with excessive force
5. Insert the QFT into the fluid source and
6. Fill the syringe with fluid using the manual fill
control.
Press the button on the tab.
7. Expel air
8. Remove QFT. Connect the disposable tubing set.
9. Follow the instructions in Section 5.5
10.
Press button to set the injection parameters
5.3.2. [Optional] IV Holder
An IV holder (part number: DC039) can be supplied as an optional
accessory. This allows IV bags and other items to be hung near the
injector.