5.7. Performing an injection
Pressing , when on the filling screen displays the current protocol screen. The routine protocol screen is
described in Section 5.7.1. More complicated multi-phase protocols may also be entered; see Section 5.8.
5.7.1. Routine Protocol Setup
The routine protocol is a simple injection protocol which delivers one phase. The volumes and injection rates
entered on this screen are not remembered by the injector when it is powered off.
To alter the default volume and flow rate values of the routine protocol, the use the Options menu from the start
screen (see Section 5.9).
[Dual Only] Saline and Contrast Routine Protocol Setup
The controls on the Dual injector screen are identical, except that a contrast phase, followed by a saline phase can
be programmed.
Alter pressure limit
Back (to manual fill screen)
Load protocol (including multi-
phase protocols)
Prime forward (0.5ml/s)
Alter injection flow rate
Alter injection volume
“Arm injector” prompt (press
orange Arm button)