5.7.2. Arming
Press the on the injector head to ARM
To disable the ARM state, press
The injector will time out of the ARM state if the injection is not started
within 10 minutes and will return to the protocol screen.
Check for air.
Press the OK button on the touch screen.
The injector is now ready to begin injecting, and the LEDs in the rear of the
head module will flash green to reflect this fact.
Tilt the head down before starting the injection to minimise risk of air
injection. With the head down, the display will reorient automatically.
5.7.3. Pressure Graph Display
Prior to starting the injection, the user can chose to
either display the syringe status or a real-time
pressure graph.
The pressure graph button is a toggle button.
- Plot not highlighted. Syringe status will be
displayed during the injection.
-
Plot highlighted. Pressure graph will be
displayed during the injection.
The selected display will remain throughout the
injection.
The pressure grap
h of the current active phase is
always displayed if a multi-
phase injection is in
progress.
Pressure Limit
Current
pressure
Pressure graph
Volume
remaining