Operation Manual

Table Of Contents
40 MEDRAD
®
Salient Operation Manual
5.8. Programming
The injector can be programmed to deliver up to 4 sequential phases of contrast injections, hold phases or delays.
The multi-phase programming mode is entered by pressing the load protocol button on the routine
protocol screen and loading any protocol except the routine protocol.
5.8.1. Multi-phase protocol setup
5.8.2. [Dual Only] Multi-phase contrast/saline protocol setup
The Dual injector multi-phase screen is similar but separate contrast and saline phases can be programmed.
Phase 2 etc
Alter phase type (see §5.8.3)
Alter pressure limit
Prime forward (0.5ml/s)
Back (to manual fill screen)
Phase 1
Load protocol (including routine
protocol)
“Arm injector” prompt (press
orange Arm button)
Save As
Alter phase volume
Alter pause time
Alter phase flow rate
Add new phase